Maintenance of HAKKO FV-310
If a failure phenomenon occurs
Maintenance of HAKKO FV-310 | |
CASE1 | Although the motor is rotating, only cool air comes out. |
CASE2 | The motor does not rotate and the heating element is not heated. |
CASE3 | Air flow volume cannot be adjusted. |
CASE 1. Although the motor is rotating, only cool air comes out.
The heating element may be broken.
The heating element deteriorates gradually during use, eventually resulting in breakage. Open the housing to check the resistance value of the heating element. * If the resistance value is abnormal, it indicates that the heating element is broken.
If the heating element is broken
Please, replace it with a new one as it is available as a replacement part.
Replacement procedure | Please refer to the instruction manual for the heating element |
Necessary tool | #1 Cross driver (150mm long is recommended) |
Part number | A5040 (110V), A5041 (220-230V), A5042 (240V) |
If the heating element is not broken
Please, send it for repair as the PWB is broken.
For repair service, please, return the product back to the store it was purchased from.